It was reported that when patient presented to clinic for normal device check, non-sustained rv oversensing episodes due to functional loss of rv capture was observed.Review of the egms shows, af with intermittent conduction, and intermittent loss of capture of the bp event, with true ventricular event occurring approximately 200ms after the bp.Programming changes were made which resolved the issue.Patient was stable and will be monitored.It was later reported that the patient expired due to congestive heart failure and coronary heart disease.The device is included in the premature battery depletion with implantable cardioverter defibrillator advisory and was explanted postmortem.
|