MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER; PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number 2AF284
Device Problem Air Leak (1008)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/12/2016
Event Type  malfunction  
Event Description
It was reported that during a cryoablation procedure, once the balloon catheter was inserted into the sheath, air aspiration was performed multiple times; however, air continuously leaked.The balloon catheter was replaced with resolve.The case was completed with cryo.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Product event summary: event data files were received and reviewed but cannot confirm the air leak seen on the date of the event.The catheter was returned, visually inspected, and functionally tested.Visual inspection of the catheter showed the device was intact with no apparent issues.Smart chip verification indicated the catheter was used for 0 injections.The catheter passed the performance test and electrical integrity as per specification; impedance was also within specification.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER
Type of Device
PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6241036
MDR Text Key64521153
Report Number3002648230-2017-00010
Device Sequence Number1
Product Code OAE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P100010/S015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 12/12/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/29/2017
Device Model Number2AF284
Device Catalogue Number2AF284
Device Lot Number21834
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2016
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/12/2016
Initial Date FDA Received01/11/2017
Supplement Dates Manufacturer ReceivedNot provided
Not provided
01/19/2017
Supplement Dates FDA Received01/26/2017
02/14/2017
09/28/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/29/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-