ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number WCD 4000
Device Problems Signal Artifact/Noise (1036); Device Sensing Problem (2917)
Patient Problem Sudden Cardiac Death (2510)
Event Date 12/14/2016
Event Type  malfunction  
Manufacturer Narrative
Device evaluations of monitor sn (b)(4) and belt sn (b)(4) are currently underway.Initial evaluation included review of downloaded software flag files on the day of the event.The review of the software flags consisted of an analysis of the downloaded data to identify any fault flags or unusual patterns of software flags.The software flag files did not suggest any device malfunction that would contribute to the patient's death.Device manufacture date: monitor: 02/21/2014, belt: 05/23/2014.
 
Event Description
A us distributor contacted zoll to report that a patient passed away on (b)(6) 2016 while wearing the lifevest.Review of the event indicates that the patient experienced an asystole event during which 2 shocks were delivered.Intermittent cardiac activity and cpr artifact contributed to the false detection.The response buttons were not pressed during the event.There is no indication that the lifevest caused or contributed to the patient's death as the patient was already in asystole, a non-life sustaining rhythm, when the shocks were delivered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
katelynn mains
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key6243243
MDR Text Key64967333
Report Number3008642652-2017-00280
Device Sequence Number1
Product Code MVK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 01/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/11/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/31/2016
Date Manufacturer Received12/16/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/30/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Death;
-
-