Model Number G148 |
Device Problems
High impedance (1291); Over-Sensing (1438); Ambient Noise Problem (2877)
|
Patient Problems
Palpitations (2467); No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/26/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further information be provided.
|
|
Event Description
|
Boston scientific received information that this cardiac resynchronization therapy defibrillator (crt-d) and another manufacturer's right atrial (ra) lead were exhibiting noise and oversensing.There was also an increase in pacing impedance values from 500 to 1,200 ohms.The respiratory rate trend was turned off.The right ventricular rate/sense and shock channels indicated appropriate ventricular activity.It was recommended by a boston scientific technical services (ts) consultant to change the sensitivity and turn on additional alerts to be notified if the signals are sensed again.No adverse patient effects were reported.
|
|
Manufacturer Narrative
|
Additional information received indicated that the patient experienced palpitations when the 112 lead ekg was taken.The local field representative performed isometrics and pocket manipulations.There was no noise or change in impedance noted.There also had been no new stored events.The physician elected to increase the ra output from 2 volts to 5 volts at 0.4 milliseconds.
|
|
Event Description
|
Additional information received indicated that loss of capture (loc) was identified when this patient was having a 12- lead electrocardiogram.Upon further review, it was identified that there was non-capture when the pacing impedance measurements were greater than 1,000 ohms, but capture returned when the pacing impedances decreased to 500 ohms.
|
|
Search Alerts/Recalls
|