MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP=-; INSULIN INFUSION PUMP / SENSOR AUGMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 640G INSULIN PUMP=-; INSULIN INFUSION PUMP / SENSOR AUGMENTED Back to Search Results
Model Number MMT-1712K
Device Problems Structural Problem (2506); Naturally Worn (2988)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/30/2016
Event Type  malfunction  
Manufacturer Narrative
The insulin pump was received with critical pump error and pump error 35 was present in long trace file due to corrosion on force sensor assembly.We were unable to perform displacement test, rewind test, prime/seating test, basic occlusion test, force sensor test, occlusion test, sleep current measurement, active current measurement and self-test due to critical pump errors.The insulin pump was received with scratched casing and minor scratches on lcd window.The insulin pump was received with pillowing keypad overlay, cracked case on battery tube area, peeling end cap address label, moisture damage on motor assembly and severe corrosion on all electronic assemblies.The insulin pump involved in this event is the 640g insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
 
Event Description
It was reported via phone call that the insulin pump's case and display window are worn and damaged.The customer's blood glucose was unknown.The customer agreed to return pump for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
640G INSULIN PUMP=-
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key6252564
MDR Text Key65250357
Report Number3004209178-2017-84036
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 01/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-1712K
Device Catalogue NumberMMT-1712K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2016
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/04/2017
Initial Date FDA Received01/14/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-