Catalog Number 400207 |
Device Problems
Device Operates Differently Than Expected (2913); Device Operational Issue (2914)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/05/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to the patient because it did not prevent the css console from performing its life-sustaining functions.The css console has been returned to syncardia for evaluation.The results of the evaluation will be provided in a follow-up mdr.(b)(4).
|
|
Event Description
|
The customer, a syncardia certified hospital, reported that the patient was being supported by the upper controller of the css console.The customer also reported that while preparing the css console lower controller for backup purposes, it could not reach the desired pressure levels before the turning knob limit was reached.There was no reported adverse patient impact.
|
|
Manufacturer Narrative
|
The css console was returned to syncardia for evaluation.The customer-reported issue was confirmed during investigation testing.The root cause was determined to be a malfunction of the pressure gauge.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4).
|
|
Event Description
|
The customer, a syncardia certified hospital, reported that the patient was being supported by the upper controller of the css console.The customer also reported that while preparing the css console lower controller for backup purposes, it could not reach the desired pressure levels before the turning knob limit was reached.There was no reported adverse patient impact.
|
|