MAUDE Adverse Event Report: SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION HOSPITAL CART

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNCARDIA SYSTEMS, LLC SYNCARDIA COMPANION HOSPITAL CART Back to Search Results
Catalog Number 397003-001
Device Problems No Display/Image (1183); Failure to Read Input Signal (1581); Device Inoperable (1663)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 01/17/2017
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to a patient because the hospital cart was not in use with a patient when the issue was observed.The companion hospital cart will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.(b)(4) initial.
 
Event Description
This syncardia companion hospital cart was not in patient use.The syncardia companion hospital cart is a large cart with wheels into which the syncardia companion 2 driver docks.It is intended for use in the hospital during the temporary total artificial heart (tah-t) implant procedure and subsequent recovery.The customer, a syncardia certified hospital, reported that the touch screen function of the hospital cart monitor did not work.
 
Manufacturer Narrative
The companion hospital cart was returned to syncardia for evaluation.The customer-reported issue of the touch screen not working was confirmed.The root cause of the touch screen not working was determined to be a malfunction of the lcd display.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.(b)(4).
 
Event Description
This syncardia companion hospital cart was not in patient use.The syncardia companion hospital cart is a large cart with wheels into which the syncardia companion 2 driver docks.It is intended for use in the hospital during the temporary total artificial heart (tah-t) implant procedure and subsequent recovery.The customer, a syncardia certified hospital, reported that the touch screen function of the hospital cart monitor did not work.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA COMPANION HOSPITAL CART
Type of Device
HOSPITAL CART
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
Manufacturer Contact
michael garippa
1992 e. silverlake road
tucson, AZ 85713
5205451234
MDR Report Key6283141
MDR Text Key66177711
Report Number3003761017-2017-00003
Device Sequence Number1
Product Code LOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/27/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number397003-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/26/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/14/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-