• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AV-TEMECULA-CT HI-TORQUE STEELCORE 18 LT GUIDE WIRE WITH MICROGLIDE COATING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT HI-TORQUE STEELCORE 18 LT GUIDE WIRE WITH MICROGLIDE COATING Back to Search Results
Catalog Number 1007710
Device Problems Difficult To Position (1467); Difficult to Remove (1528); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/04/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was received.Investigation is not yet complete.A follow up report will be submitted with all additional relevant information.The supera referenced is being filed under a separate medwatch report number.
 
Event Description
It was reported that the procedure was to treat a moderately calcified superficial femoral artery.During deployment of the 5.5 x 150 mm supera self-expanding stent, it was noted that the deployment felt different.The stent was fully deployed but when trying to advance the steel core guide wire, it froze in the supera delivery system and the tip separated.The guide wire was wrapped around the delivery system when removed from the anatomy and the tip was on the guide wire.The procedure was completed at this time.There was no clinically significant delay in procedure and no adverse patient effects.No additional information was provided.
 
Manufacturer Narrative
Internal file number - (b)(4).Evaluation summary: visual and dimensional inspections were performed on the returned device.The reported difficulties could not be confirmed as it was based on operational circumstances of the procedure.The investigation determined the reported difficulties to be related to the patient anatomical conditions.A review of the lot history record and complaint history of the reported lot could not be conducted because the lot number was not provided.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HI-TORQUE STEELCORE 18 LT GUIDE WIRE WITH MICROGLIDE COATING
Type of Device
GUIDE WIRE
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, TEMECULA, CA USA REG# 2024168
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6283903
MDR Text Key66147249
Report Number2024168-2017-00740
Device Sequence Number1
Product Code DQX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K980119
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 02/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number1007710
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer01/16/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/04/2017
Initial Date FDA Received01/27/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received02/07/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
STENT: SUPERA
Patient Age37 YR
Patient Weight98
-
-