• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED Back to Search Results
Model Number MMT-754CAH
Device Problems Dent in Material (2526); Naturally Worn (2988)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/05/2017
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.The insulin pump involved in this event is the paradigm real-time veo insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
 
Event Description
The customer reported via phone call that the bottom of the reservoir compartment was worn out and the drive support cap was indented.The customer's blood glucose was 10.9 mmol/l at the time of incident.The customer stated that the drive support cap was flush.The customer was advised that the insulin pump will need to be replaced.The customer was advised to revert to back-up plan.The insulin pump will not be returned for analysis.
 
Manufacturer Narrative
The insulin pump was received with partially detached end cap.Drive support disk was inspected and no anomaly was noted.Device was received with cracked case at display window corner, cracked lcd window, minor scratched lcd window, cracked battery tube threads and cracked reservoir tube lip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME INSULIN INFUSION PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key6284542
MDR Text Key66283610
Report Number3004209178-2017-87366
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P980022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-754CAH
Device Catalogue NumberMMT-754CAH
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/09/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/05/2017
Initial Date FDA Received01/27/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/24/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-