• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUADRA ASSURA (DDQP+); IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) QUADRA ASSURA (DDQP+); IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD3265-40Q
Device Problem Vibration (1674)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/10/2017
Event Type  malfunction  
Manufacturer Narrative
Udi (di): (b)(4).
 
Event Description
It was reported that the patient cannot feel the vibratory notifier when tested in clinic.After two attempts the troubleshooting of the vibration notifier has been exhausted.The decision was made to replace the device in (b)(6) 2017.
 
Event Description
New information notes that the device was explanted and replaced.The device is included in the premature battery depletion with implantable cardioverter defibrillator advisory.No issue was reported regarding premature eri.Patient is a pacemaker dependent.
 
Manufacturer Narrative
No conclusion code available.Final analysis found the reported field event of patient notifier anomaly was confirmed in the laboratory.The notifier was tested in the lab and was found to be anomalous.The device was tested on the bench and no anomalies were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA (DDQP+)
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6294880
MDR Text Key66488864
Report Number2938836-2017-13425
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 11/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/31/2014
Device Model NumberCD3265-40Q
Device Catalogue NumberCD3265-40Q
Device Lot Number4133609
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/16/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/07/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/25/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age77 YR
-
-