• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES PR SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER; SWAN-GANZ CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES PR SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER; SWAN-GANZ CATHETER Back to Search Results
Model Number 774F75
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/16/2016
Event Type  malfunction  
Manufacturer Narrative
We received one 774f75 catheter with 1.5 cc monoject limited volume syringe and cal-cup for examination.The reported event of "balloon was not able to inflate" was confirmed.The balloon was torn at both the proximal and distal bonds.Most of the balloon was missing and not returned.Adhesive residue and leftover latex was still attached to the proximal and distal bonds were found.All through lumens were patent without any leakage or occlusion.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint.A review of the manufacturing records indicated that the product met specifications upon release.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised, before deciding to insert or use the catheter, to consider the potential benefits in relation to the possible complications.The techniques for insertion, methods of using the catheter to obtain patient data information, and the occurrence of complications is well described in the literature.It is common clinical practice to check balloon integrity by inflating it to the recommended volume in order to detect any asymmetry or leakage condition before use of the catheter.When there is separation of the balloon or fragments from the pulmonary artery catheter, the retained fragment will embolize to the lungs.Due to the large surface area of the pulmonary vasculature, this is generally well tolerated.Pulmonary complications may result from improper inflation technique.To avoid damage to the pulmonary artery and possible balloon rupture, the balloon should not be inflated above the recommended volume.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
The balloon was not able to inflate during testing.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER
Type of Device
SWAN-GANZ CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
brian hurd-servin
one edwards way
irvine, CA 92614
9492506423
MDR Report Key6296399
MDR Text Key66360708
Report Number2015691-2017-00240
Device Sequence Number1
Product Code DQE
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K040287
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Physician
Type of Report Initial
Report Date 12/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/11/2018
Device Model Number774F75
Device Lot Number60335040
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/10/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/11/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/14/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-