Model Number MMT-523NAL |
Device Problems
Device Displays Incorrect Message (2591); Electromagnetic Compatibility Problem (2927)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/27/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
|
Event Description
|
The customer reported via phone call that the insulin pump was exposed to an mri and it cleared the software.Customer's blood glucose level was not reported.The insulin pump alarmed motor error and motor position encoder error.Customer was advised to discontinue use of the device and revert to backup plan.The customer was advised that the device would be replaced and agreed to return the product for analysis.
|
|
Manufacturer Narrative
|
The insulin pump alarmed motor error during rewind due to motor encoder signal out of phase.Unable to perform displacement test due to motor anomaly.Unable to confirm motor position encoder error alarm due to erased history file.No cosmetic damage noted.
|
|