Model Number MMT-523NAL
Device Problems Device Displays Incorrect Message (2591); Electromagnetic Compatibility Problem (2927)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/27/2017
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that the insulin pump was exposed to an mri and it cleared the software.Customer's blood glucose level was not reported.The insulin pump alarmed motor error and motor position encoder error.Customer was advised to discontinue use of the device and revert to backup plan.The customer was advised that the device would be replaced and agreed to return the product for analysis.
 
Manufacturer Narrative
The insulin pump alarmed motor error during rewind due to motor encoder signal out of phase.Unable to perform displacement test due to motor anomaly.Unable to confirm motor position encoder error alarm due to erased history file.No cosmetic damage noted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key6297214
MDR Text Key66793434
Report Number3004209178-2017-88441
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
PMA/PMN Number
P980022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 02/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/02/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-523NAL
Device Catalogue NumberMMT-523NAL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received02/25/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age40 YR