• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - CALI MINICAP; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - CALI MINICAP; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE Back to Search Results
Catalog Number ARM4466E
Device Problem Unsealed Device Packaging (1444)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).The device was returned and an evaluation is complete.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.A visual inspection was performed and the lower seal was not complete and seal separation was evidenced.The reported condition was verified and the cause is attributed to improper positioning of the manufacturing machine by the operator.A nonconformance has been opened to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the foil packaging for a minicap was received without air.This was observed before use.There was no patient involvement.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINICAP
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE - CALI
cali
Manufacturer (Section G)
BAXTER HEALTHCARE - CALI
calle 36 no. 2c-22,
apartado aero 2446
cali
CO  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key6299934
MDR Text Key66459112
Report Number1416980-2017-00865
Device Sequence Number1
Product Code KDJ
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial
Report Date 02/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2018
Device Catalogue NumberARM4466E
Device Lot NumberSE16IB4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/30/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/11/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-