MAUDE Adverse Event Report: SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY AXIOM ARTIS BA; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY AXIOM ARTIS BA; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Model Number 5904656
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problems No Known Impact Or Consequence To Patient (2692); Shock from Patient Lead(s) (3162)
Event Date 01/05/2017
Event Type  malfunction  
Manufacturer Narrative
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.
 
Event Description
It was reported to siemens that a malfunction occurred while operating the axiom artis ba system.The customer reported receiving a shock after the table control module (tcm) failed during a procedure.We are unaware of any impact to the state of health of any patient or user involved.
 
Manufacturer Narrative
Siemens has completed an investigation of the reported event.Assessment of the log files does not indicate a system failure or malfunction and no non-conformity was identified.Tests on site and tests in the lab of the components in question passed without any problems.In this case, an electrical shock is technically not possible as the power at the control module is on a low range of 24v only.The table control module (tcm) and the computer area network (can) terminator plug have been exchanged as a preventative action.The manufacturer is not considering further actions resulting from this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIOM ARTIS BA
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
SIEMENS HEALTHCARE, BUSINESS AREA ADVANCED THERAPY
siemensstrasse 1
forcheim, 91301
GM  91301
Manufacturer (Section G)
SIEMENS HEALTHCARE GMBH
siemenstrasse 1
forcheim, 91301
GM   91301
Manufacturer Contact
meredith adams
40 liberty blvd.
65-1a
malvern, PA 19355
6104486461
MDR Report Key6300823
MDR Text Key66770240
Report Number2240869-2017-66439
Device Sequence Number1
Product Code OWB
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K052202
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Remedial Action Inspection
Type of Report Initial,Followup
Report Date 01/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number5904656
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received12/05/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/07/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-