• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES, PR SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER WITH AMC THROMBOSHIELD; SWAN-GANZ CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES, PR SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER WITH AMC THROMBOSHIELD; SWAN-GANZ CATHETER Back to Search Results
Model Number 774HF75J
Device Problem Material Rupture (1546)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/06/2016
Event Type  malfunction  
Manufacturer Narrative
One catheter with attached monoject 1.5 cc limited volume syringe and a three-way stopcock were returned for evaluation.A non-edwards contamination shield was returned with the unit.No introducer or package was returned.Balloon was found to be ruptured around the central area of the balloon latex.The ruptured edge of the latex appeared to be different shapes and was not able to match up.All through lumens were patent without any leakage or occlusion.No other visible damage or inconsistency to the catheter body, balloon, or returned syringe was observed.Visual examination was performed under microscope at 20x magnification and with the unaided eyes.A device history record review was completed and documented that device met all specifications upon distribution.The customer report of balloon inflation issue was confirmed.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint and implement any necessary corrective actions.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised, before deciding to insert or use the catheter, to consider the potential benefits in relation to the possible complications.The techniques for insertion, methods of using the catheter to obtain patient data information, and the occurrence of complications is well described in the literature.It is common clinical practice to check balloon integrity by inflating it to the recommended volume in order to detect any asymmetry or leakage condition before use of the catheter.When there is separation of the balloon or fragments from the pulmonary artery catheter, the retained fragment will embolize to the lungs.Due to the large surface area of the pulmonary vasculature, this is generally well tolerated.Pulmonary complications may result from improper inflation technique.To avoid damage to the pulmonary artery and possible balloon rupture, the balloon should not be inflated above the recommended volume.It is unknown if user or procedural factors may have contributed to the event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
 
Event Description
It was reported that the balloon could not maintain its inflation during use.There were no patient complications reported.Inquired of patient demographics.Unable to be obtained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWAN-GANZ CCOMBO V CCO/SVO2/CEDV THERMODILUTION CATHETER WITH AMC THROMBOSHIELD
Type of Device
SWAN-GANZ CATHETER
Manufacturer (Section D)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES, PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
lynn selawski
1 edwards way
irvine, CA 92614
9497564386
MDR Report Key6301868
MDR Text Key66553120
Report Number2015691-2017-00267
Device Sequence Number1
Product Code DQE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K040287
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 12/06/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/08/2017
Device Model Number774HF75J
Device Lot Number60359572
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/04/2017
Initial Date Manufacturer Received 01/11/2016
Initial Date FDA Received02/03/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/11/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-