(b)(4).The device was returned for evaluation and visually and functionally tested pursuant to (b)(4).Upon inspection it was found that the electrodes on the active and inactive side were pulled from the device.This particular lot was identified as one that was built in the time frame in which rev 05 electrodes were being used in production.Electrode dislodgement has been investigated per (b)(4) and concluded that all electrode dislodgement complaints have been from devices built with rev 05 electrodes.Per (b)(4), the electrodes were updated, implemented into production, and as a result, since this complaint, the electrode dislodgement issue has since been mitigated.
|
During a mv repair, cabg, and pv isolation procedure, sections 1 and 2 on the probe were faulty.All other sections functioned well and as expected.The patient was off pump and outcome was not affected.
|