• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW JUNIOR INTERFACE; CAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD OPTIFLOW JUNIOR INTERFACE; CAT Back to Search Results
Model Number OPT318
Device Problem Device Handling Problem (3265)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/13/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Method: the complaint opt318 cannula has to date not been received for evaluation.Our investigation is based on photographs provided by the hospital and our knowledge of the product.Results: visual inspection of the photographs revealed that one of the flexitubes (cannula tubing) was damaged at the swivel grip joint, which connects to the breathing circuit.The tubing was stretched and broken but the metal spring was still attached to the swivel grip.Conclusion: the damage to the flexitube is likely to have been caused by excessive pulling force being exerted on the tubing.We were informed that the patient was (b)(6) and "very active." all optiflow junior cannulae are 100% leak and occlusion tested after final assembly and any cannula that fails is discarded.In addition, samples are taken hourly from each run and pull tested to check glue joint strength at the cannula/tube joint, as well as the swivel grip joint.If there are any failures the entire batch is put aside for further investigation.The user instructions illustrate in pictorial format the correct set-up and proper use of the optiflow junior nasal cannula.They also state the following: ensure that all connections are secure during use.Check cannula is undamaged and that the flow path is maintained.Appropriate monitoring must be used at all times.Do not stretch or crush tube.
 
Event Description
A hospital in (b)(6) reported that the tubing of an opt318 optiflow junior nasal cannula used on a (b)(6) patient was "disconnected from the adaptor that connects to the heated breathing tube." no patient consequence was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLOW JUNIOR INTERFACE
Type of Device
CAT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
raymond yan
173 technology drive
suite 100
irvine, CA 92618
9494534000
MDR Report Key6318965
MDR Text Key67231624
Report Number9611451-2017-00086
Device Sequence Number1
Product Code CAT
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Respiratory Therapist
Type of Report Initial
Report Date 01/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOPT318
Device Catalogue NumberOPT318
Device Lot Number2100065328
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/13/2017
Initial Date FDA Received02/09/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/20/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age17 MO
-
-