The customer contacted a siemens customer care center.The customer treated the patient sample with a heterophilic blocking tube and tested it on the immulite 2000 xpi instrument, using reagent kit lot 410 and the results were lower.This indicated that heterophilic antibodies were elevating the result of this sample with the immulite 2000 xpi psa assay.As stated in the immulite 2000 psa instructions for use (ifu), "heterophilic antibodies in human serum can react with the immunoglobulins included in the assay components causing interference with in vitro immunoassays.Samples from patients routinely exposed to animals or animal serum products can demonstrate this type of interference potentially causing an anomalous result.These reagents have been formulated to minimize the risk of interference; however, potential interactions between rare sera and test components can occur.For diagnostic purposes, the results obtained from this assay should always be used in combination with the clinical examination, patient medical history, and other findings." a siemens headquarters support center specialist reviewed the sample data and determined that immulite 2000 psa reagent kit lot 410 is performing as intended.No further evaluation of device is required.
|
Discordant, falsely elevated prostate specific antigen (psa) results were obtained on one patient on an immulite 2000 xpi instrument, when using reagent kit lot 410.The sample was initially tested on an alternate platform, where the result was lower.The sample was then re-run after treatment with heterophilic blocking tube on both the immulite 2000 xpi instrument and on the alternate platform.The results obtained after treatment were lower on the immulite 2000 xpi platform.The discordant results were not reported to the physician(s).The correct results obtained on the alternate platform were reported to the physician(s).There are no known reports of patient intervention or adverse health consequence as a result of discordant, falsely elevated psa results.
|