• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN KANGAROO JOEY,PUMP W/POLE CLMP; PUMP, INFUSION, ENTERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN KANGAROO JOEY,PUMP W/POLE CLMP; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number 383400
Device Problems Device Stops Intermittently (1599); Failure to Deliver (2338)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Submit date: 02/13/2017.An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event Description
It was reported to covidien on (b)(6) 2017 that an issue occurred an enteral feeding pump.On (b)(6) 2017 the customer clarified that the unit stops working during feeds, sometimes it won't work at all.There was no medical intervention.
 
Manufacturer Narrative
The unit was triaged and the customer¿s reported condition was confirmed.The root cause of the reported condition was due to the battery¿s inability to hold a charge.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO JOEY,PUMP W/POLE CLMP
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN
building # 10, 789 puxing road
shanghai
CH 
Manufacturer (Section G)
COVIDIEN
building # 10, 789 puxing road
shanghai
CH  
Manufacturer Contact
edward almeida
15 hampshire st
mansfield, MA 02048
5084524151
MDR Report Key6326657
MDR Text Key67485295
Report Number3006451981-2017-00176
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/13/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number383400
Device Catalogue Number383400
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received02/10/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-