• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLOGIC, MARLBORUGH MA MYOSURE HYSTEROSCOPIC TISSUE REMOVAL SYSTEM; UTERINE TISSUE REMOVAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, MARLBORUGH MA MYOSURE HYSTEROSCOPIC TISSUE REMOVAL SYSTEM; UTERINE TISSUE REMOVAL SYSTEM Back to Search Results
Catalog Number 10-401
Device Problems Break (1069); Detachment Of Device Component (1104)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Device history record (dhr) review was conducted for the reported identification number.The lot was released meeting all qa specifications.Currently unable to establish a relationship or impact to the reported observation.(b)(4).
 
Event Description
It was reported during a myosure procedure for uterine tissue removal (exact date unknown), the physician noticed "the end of the myosure [disposable device] broke off while in use." a second device completed the procedure.There was no patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MYOSURE HYSTEROSCOPIC TISSUE REMOVAL SYSTEM
Type of Device
UTERINE TISSUE REMOVAL SYSTEM
Manufacturer (Section D)
HOLOGIC, MARLBORUGH MA
250 campus drive
marlborough MA 01752
Manufacturer Contact
sidra piracha
250 campus drive
marlborough, MA 01752
5082638884
MDR Report Key6327397
MDR Text Key67581916
Report Number1222780-2017-00028
Device Sequence Number1
Product Code HIH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K100559
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date09/19/2017
Device Catalogue Number10-401
Device Lot Number16B16RK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/03/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/17/2017
Initial Date FDA Received02/13/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/07/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MYOSURE CONTROL UNIT-SERIAL # UNK; MYOSURE HYSTEROSCOPE-SERIAL # UNK
-
-