Device evaluation of electrode belt sn (b)(4) has been completed.The reported problem (delivered pulse below lower limit) has been investigated.As received, the belt failed incoming functionality testing, specifically the therapy electrode recognition test.Upon investigation the electrode belt had an open pulse wire between the front therapy electrode and ecg 'a'.The cause of the test failure is the open pulse wire.The root cause of the open pulse wire is excessive force on the cable.No adverse event resulted from the defective electrode belt.
|