• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARIBEX NOMAD PRO 2; EXTRAORAL SOURCE X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARIBEX NOMAD PRO 2; EXTRAORAL SOURCE X-RAY SYSTEM Back to Search Results
Model Number 0.850.0021
Device Problem Battery Problem (2885)
Patient Problem No Patient Involvement (2645)
Event Date 01/13/2017
Event Type  malfunction  
Manufacturer Narrative
Unit has been returned and is undergoing evaluation.A follow-up report will be submitted upon evaluation completion.A recall has been ongoing, refer to z-2716/2717-2016.
 
Event Description
It was initially reported by the dealer representative that the handset caught fire.The dental office reported that they had smelled smoke when they came into the office and saw brown marks on the cabinet.It was reported the unit was not in use at the time of the event.There was no patient or user involvement and no injuries reported.There was no reported impact to patient care at the time of the event.
 
Manufacturer Narrative
Upon visual inspection, it appears that a thermal event occurred.It is not possible to determine the exact sequence of events that led to the thermal event.However, there is evidence that a short occured in the lower battery pack between the cells and pcb.There was a blackening on the top of the lower battery pack and its printed circuit board (pcb).Cells 2 and 3 appear to have been involved in the thermal event, likely by supplying energy into a fault in the board.Additionally, there was a slight melting of the handset plastic enclosure.This concludes the investigation.A recall is ongoing.Reference z-2716/2717-2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOMAD PRO 2
Type of Device
EXTRAORAL SOURCE X-RAY SYSTEM
Manufacturer (Section D)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer (Section G)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer Contact
erika martin
11727 fruehauf drive
charlotte, NC 28273
7045877241
MDR Report Key6341515
MDR Text Key67821611
Report Number1017522-2017-00006
Device Sequence Number1
Product Code EHD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081664
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0.850.0021
Device Catalogue Number0.850.0021
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/23/2017
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 01/18/2017
Initial Date FDA Received02/17/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/01/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/08/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-