• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 40 PARKER RADICAL 7; OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 40 PARKER RADICAL 7; OXIMETER Back to Search Results
Model Number 23789
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/30/2017
Event Type  malfunction  
Manufacturer Narrative
Additional manufacuring narrative: attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.(b)(6).
 
Event Description
It was reported that "display changed by itself without touching and unable to change the brightness by touching".No known impact or consequence to patient.
 
Manufacturer Narrative
The returned device was evaluated.During evaluation, the unit was able to turn on; however, the screen was scrolling by itself.It was determined that the touchscreen was defective, causing the screen to change without being touched.As a result, the device was not able to engage in active monitoring.A service history record review reveals that this unit was in the field for over four (4) years with no previous reported issues prior to this reported event.The customer zip code exceeded maximum allowable digits, zip code is as follows: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RADICAL 7
Type of Device
OXIMETER
Manufacturer (Section D)
MASIMO - 40 PARKER
40 parker
irvine CA 92618 1604
Manufacturer (Section G)
INDUSTRIAL VALLERA DE MEXICALI S.A DE C.V.
calzada del oro, no. 2001
parque industrial palaco
mexicali, baja california 21600
MX   21600
Manufacturer Contact
erica kline
52 discovery
irvine, CA 92618-1604
9492977000
MDR Report Key6344463
MDR Text Key67823690
Report Number2031172-2017-00233
Device Sequence Number1
Product Code DQA
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K140188
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Biomedical Engineer
Remedial Action Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number23789
Device Catalogue Number9500
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Initial Date Manufacturer Received 01/30/2017
Initial Date FDA Received02/20/2017
Supplement Dates Manufacturer Received03/12/2018
Supplement Dates FDA Received04/09/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/25/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-