Model Number CD3251-40Q
Device Problem Telemetry Discrepancy (1629)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/23/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
A merlin transmission was received which noted the rv and lv lead¿s output to be increased, though no programming changes had been completed.Upon further testing, the output of the rv and lv lead was within normal limits.An additional follow up was done and again confirmed the output to be normal.It is unknown why there was a discrepancy between the transmission measurement and the clinic measurement.There were no adverse consequences to the patient.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.The issue was resolved with reprogramming.
 
Manufacturer Narrative
Additional information: the results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
During subsequent follow up, the patient returned to the clinic and the ac current was disabled and fixed outputs were programmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY QUADRA CRT-D, OUS MODEL, DF4COMP
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6347233
MDR Text Key68176555
Report Number2938836-2017-15749
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 01/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/28/2013
Device Model NumberCD3251-40Q
Device Catalogue NumberCD3251-40Q
Device Lot Number3484530
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/23/2017
Initial Date FDA Received02/21/2017
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received03/20/2017
03/23/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/26/2011
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-