Model Number CD3371-40C
Device Problem Pacing Inadequately (1442)
Patient Problem Ventricular Fibrillation (2130)
Event Date 01/14/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.Udi: (b)(4).
 
Event Description
The patient experienced ventricular fibrillation due to unsuccessful anti-tachycardia pacing.The device delivered four shocks before the rhythm converted to sinus rhythm.A subcutaneous coil was implanted in addition to the existing system to resolve the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA MP ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6349731
MDR Text Key68031407
Report Number2938836-2017-15905
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeDA
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 01/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/22/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2018
Device Model NumberCD3371-40C
Device Lot NumberP000028823
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/23/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/05/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-