• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ERIKA DE REYNOSA, S.A. DE C.V. LIBERTY CYCLER SET, SINGLE CONN./EXT. DL; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ERIKA DE REYNOSA, S.A. DE C.V. LIBERTY CYCLER SET, SINGLE CONN./EXT. DL; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 050-87216
Device Problems Fluid/Blood Leak (1250); Hole In Material (1293)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/27/2017
Event Type  malfunction  
Manufacturer Narrative
Unique identifier (udi): (b)(4) - the set was returned for evaluation.A visual inspection was performed and a hole was noted on the cassette membrane.No damage on the hard plastic was noted.An investigation of the device manufacturing records was conducted by the manufacturer.There were no deviations or non-conformances during the manufacturing process.In addition, the batch record review confirmed the labeling, material, and process controls were within specification.
 
Event Description
A peritoneal dialysis registered nurse reported that the patient went in for treatment at the clinic.Following treatment when they removed the cassette the inside of the door was wet.There were no alarms and the patient was able to complete the treatment.Set was made available for evaluation.During follow up the patient's pd nurse reported that the patient did not have any signs of infection.Her effluent remained clear.She was not prescribed any antibiotics.No adverse event reported at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIBERTY CYCLER SET, SINGLE CONN./EXT. DL
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX  88780
Manufacturer (Section G)
ERIKA DE REYNOSA, S.A. DE C.V.
mike allen #1331
parque industrial reynosa
reynosa 88780
MX   88780
Manufacturer Contact
thomas c. johnson
920 winter st.
waltham, MA 02451
7816999499
MDR Report Key6357130
MDR Text Key68382097
Report Number8030665-2017-00074
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K043363
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Nurse
Type of Report Initial
Report Date 02/22/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/24/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/31/2019
Device Catalogue Number050-87216
Device Lot Number16NR08013
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2017
Is the Reporter a Health Professional? Yes
Device AgeMO
Event Location Dialysis Center
Date Manufacturer Received01/27/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/12/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
LIBERTY CYCLER; PD FLUID
Patient Age67 YR
Patient Weight102
-
-