• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MASIMO - 40 PARKER RAD-8; OXIMETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MASIMO - 40 PARKER RAD-8; OXIMETER Back to Search Results
Model Number 20877
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/03/2017
Event Type  malfunction  
Manufacturer Narrative
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.(b)(6).
 
Event Description
It was reported that after locking the setting in on the device, the setting will not stick.Per customer, "what we were wanting to do was change the parameters on the monitor that were more appropriate to a neonate in the birth suite at birth.Currently the monitor is changed prior to being used at a birth to reflect appropriate levels for a newborn but this has to be repeated each time as it returns to the original settings.If this isn't done prior to being used at a birth it means during a resuscitation the monitor constantly alarms for levels that are tolerated on a baby immediately following birth." no known impact or consequence to patient.
 
Manufacturer Narrative
The returned device was evaluated.During evaluation the device passed all visual and functional testing.The unit was found to visually and audibly alarm during alarm conditions.The unit was determined to be functioning as designed.Customer address exceeded maximum allowable characters, address is as follows: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAD-8
Type of Device
OXIMETER
Manufacturer (Section D)
MASIMO - 40 PARKER
40 parker
irvine CA 92618 1604
Manufacturer (Section G)
INDUSTRIAL VALLERA DE MEXICALI S.A DE C.V.
calzada del oro, no. 2001
parque industrial palaco
mexicali, baja california 21600
MX   21600
Manufacturer Contact
charlene johnson
52 discovery
irvine, CA 92618-1604
9492977000
MDR Report Key6359173
MDR Text Key68356894
Report Number2031172-2017-00257
Device Sequence Number1
Product Code DQA
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
K092838
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/24/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20877
Device Catalogue Number9019
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Date Manufacturer Received03/06/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/03/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-