• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS 3MM FLTD MATCHSTK,EXT 8.9MM,LESS AGGRESS; DRILLS, BURRS, TREPHINES AND ACCESSORIES (SIMPLE, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS 3MM FLTD MATCHSTK,EXT 8.9MM,LESS AGGRESS; DRILLS, BURRS, TREPHINES AND ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Catalog Number S-8NS-M
Device Problem Difficult To Position (1467)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/21/2016
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that during a craniotomy surgical procedure, it was observed that the cutter device would not fit into the attachment device.There were no delays to the surgical procedure as a spare device was available for use.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
(b)(4).The device manufacture date was documented as unknown in the initial report.It has been updated as oct 6, 2015.The actual device was returned for evaluation.Reliability engineering evaluated the device and observed that the device passed all operational specifications.Therefore, the reported condition was not confirmed.An assignable root cause was not determined.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3MM FLTD MATCHSTK,EXT 8.9MM,LESS AGGRESS
Type of Device
DRILLS, BURRS, TREPHINES AND ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key6360113
MDR Text Key68360468
Report Number1045834-2017-10481
Device Sequence Number1
Product Code HBE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113476
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/06/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberS-8NS-M
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received04/24/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/06/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
ATTACHMENT DEVICE
-
-