(b)(4).The opt870 interface is used to deliver humidified oxygen to patients via tracheostomy.The interface is held in place by a neck strap and also includes a lanyard which is placed around the patient's neck or attached to the patient's clothing or bedding to remove the load of the breathing circuit from the patient's trache.The complaint opt870 was not returned to fisher & paykel healthcare (b)(4) for evaluation.Further information was sought from the customer but to date no response has been received.Conclusion: the customer had informed us that the opt870 trache interface disconnected where the tubing meets the connector.Based on this information, it appears likely that the tubing had become disconnected from the swivel that connects to the breathing tube.Without a device for evaluation we cannot determine what has caused the device to malfunction in this way.This is the only complaint of this nature that we have received for the opt870.All optiflow interfaces are inspected during production for visual defects including cracks, tears, inclusions, discolouration and deformation.Any product that fails the visual inspection is rejected.Our user instructions that accompany the opt870 trache interface state: to ensure loading on tracheostomy tube is kept to a minimum, make sure lanyard is fitted appropriately.Do not crush or stretch tube.
|