FISHER & PAYKEL HEALTHCARE LTD ADULT TRACHE DIRECT INTERFACE; BZA Back to Search Results
Model Number OPT870
Device Problem Connection Problem (2900)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The opt870 interface is used to deliver humidified oxygen to patients via tracheostomy.The interface is held in place by a neck strap and also includes a lanyard which is placed around the patient's neck or attached to the patient's clothing or bedding to remove the load of the breathing circuit from the patient's trache.The complaint opt870 was not returned to fisher & paykel healthcare (b)(4) for evaluation.Further information was sought from the customer but to date no response has been received.Conclusion: the customer had informed us that the opt870 trache interface disconnected where the tubing meets the connector.Based on this information, it appears likely that the tubing had become disconnected from the swivel that connects to the breathing tube.Without a device for evaluation we cannot determine what has caused the device to malfunction in this way.This is the only complaint of this nature that we have received for the opt870.All optiflow interfaces are inspected during production for visual defects including cracks, tears, inclusions, discolouration and deformation.Any product that fails the visual inspection is rejected.Our user instructions that accompany the opt870 trache interface state: to ensure loading on tracheostomy tube is kept to a minimum, make sure lanyard is fitted appropriately.Do not crush or stretch tube.
 
Event Description
A hospital in (b)(6) reported that the opt870 tracheostomy interface "often disconnects." no patient consequence was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADULT TRACHE DIRECT INTERFACE
Type of Device
BZA
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
jon stevens
15365 barranca parkway
irvine, CA 92618-2216
9194534000
MDR Report Key6362759
MDR Text Key68470770
Report Number9611451-2017-00137
Device Sequence Number1
Product Code BZA
Combination Product (y/n)N
Reporter Country CodeNZ
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other
Type of Report Initial
Report Date 02/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOPT870
Device Catalogue NumberOPT870
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/01/2017
Initial Date FDA Received02/27/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-