STRYKER ORTHOPAEDICS-MAHWAH CUP IMPACTOR BOLT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 1440-2011
Device Problems Mechanical Problem (1384); Difficult To Position (1467); Sticking (1597); Mechanical Jam (2983); Positioning Problem (3009)
Patient Problems No Consequences Or Impact To Patient (2199); No Information (3190)
Event Date 01/31/2017
Event Type  malfunction  
Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event Description
Impactor bolts stuck in cups.
 
Manufacturer Narrative
An event regarding a disassembly issue for a cup impactor bolt from an associated shell was reported.The event was confirmed.Visual inspection: the bolt was returned assembled with the trident shell.The device was visually unremarkable.A capa investigation was issued for exceeded complaint rate and concluded insufficient instructions/training as root cause of these events.Medical records received and evaluation: not performed as patient factors did not contribute to the reported event.Device history review: all devices in the reported lot were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been two other events for the lot reported.The event was confirmed.
 
Event Description
Impactor bolts stuck in cups.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUP IMPACTOR BOLT
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
Manufacturer Contact
keyla colon
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key6366464
MDR Text Key69014346
Report Number0002249697-2017-00696
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K153345
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number1440-2011
Device Lot NumberER9ED6
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/22/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/14/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/14/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age63 YR
-
-