The account found the brake detent assembly broken.Hill-rom technical support supplied the account with the part number for the brake detent assembly.Per the hill-rom service manual the affinity® three birthing bed and affinity® four birthing beds require an effective maintenance program.We recommend that you perform semiannual preventive maintenance.Check the tires for cuts, wear, tread life, etc.Apply the brake, and check to ensure that the bed will not move.If the bed moves, inspect it for wear, and adjust if required.Apply the steering pedal and check the steering to ensure proper locking action when activated.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The account will replace the brake detent assembly to resolve the issue.Based on this information, no further action is required.
|