Model Number 07555357 |
Device Problems
Device Emits Odor (1425); Noise, Audible (3273)
|
Patient Problem
No Information (3190)
|
Event Date 01/31/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Siemens is conducting a thorough investigation of the reported events.As this event is under investigation, a root cause has not yet been determined.A supplement report will be filed upon completion of the investigation.This event occurred in (b)(6).
|
|
Event Description
|
It was reported to siemens that a malfunction occurred after use of the axiom artis dba system.Following an examination, the customer reported no image was displayed on the monitor and an explosive sound and odor from the isolation amplifier vga.We are unaware of any impact to the state of health of any patient involved.
|
|
Manufacturer Narrative
|
Siemens has completed an investigation of the report.During the investigation, it was determined that the primary circuit of the power supply of the isolation amplifier otv was defective caused by excessive moisture inside the unit.Moisture traces and deposits have been found inside the housing and on the circuit board.The isolation amplifier otv-vga-v2 has been exchanged and system works as specified.The manufacturer is not considering further actions resulting from this event.
|
|