Model Number P3200K000037
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/17/2017
Event Type  malfunction  
Manufacturer Narrative
The hill-rom technician found the scale board needed to be replaced.Per the hill-rom user manual, warning: the bed exit alarm system is not intended as a substitute for good nursing practices.The bed exit alarm system must be used in conjunction with a sound risk assessment and protocol.Per the hill-rom user manual, if the bed exit alarm does not arm and all three mode indicators are flashing, remove the patient and zero the bed exit system.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the scale board to resolve the issue.Based on this information, no further action is required.
 
Event Description
Hill-rom received a report from the account stating that only one part of the ppm was functioning.The bed was located at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACARE FRAME
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM BATESVILLE
1069 state route 46 east
batesville IN 47006
Manufacturer Contact
erin padgett
1069 state route 46 east
batesville, IN 47006
8129312791
MDR Report Key6377079
MDR Text Key69285540
Report Number1824206-2017-00106
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 02/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Caregivers
Device Model NumberP3200K000037
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/17/2017
Initial Date FDA Received03/03/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/14/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-