• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARIBEX NOMAD PRO 2; EXTRAORAL SOURCE X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARIBEX NOMAD PRO 2; EXTRAORAL SOURCE X-RAY SYSTEM Back to Search Results
Model Number 0.850.0021
Device Problem Battery Problem (2885)
Patient Problem No Patient Involvement (2645)
Event Date 12/01/2016
Event Type  malfunction  
Manufacturer Narrative
Upon visual inspection it appears that a thermal event occurred.It is not possible to determine the exact sequence of events that led to the thermal event.However, there is evidence that a short occurred in the upper battery pack between the cells and printed circuit board (pcb).There was a blackening on the top of the upper battery pack and its printed circuit board (pcb).Cells 5 and 6 appear to have been involved in the thermal event, likely by supplying energy into a fault in the board.Additionally there was a slight melting of the handset plastic enclosure.This concludes the investigation.Upon initial receipt of the complaint, aribex did not believe the event was a reportable event.However, during an investigation on 02/06/2017, it was determined that the event is reportable.A recall is ongoing.Reference z-2716/2717-2016.
 
Event Description
It was reported that the battery is giving red lights on the charger.The doctor noticed a burn mark on the inside of the cabinet where it looks like the battery exploded and is in the shape of the battery handset.No injuries, user or patient involvement were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOMAD PRO 2
Type of Device
EXTRAORAL SOURCE X-RAY SYSTEM
Manufacturer (Section D)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer (Section G)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer Contact
erika martin
11727 fruehauf drive
charlotte, NC 28273
7045877240
MDR Report Key6386346
MDR Text Key69329393
Report Number1017522-2017-00014
Device Sequence Number1
Product Code EHD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081664
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 12/01/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/07/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0.850.0021
Device Catalogue Number0.850.0021
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2016
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/01/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/08/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-