Model Number 3788
Device Problem Use of Device Problem (1670)
Patient Problem Inadequate Pain Relief (2388)
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported the patient is unable to establish communication between the ipg and external devices.The programmer also reflects a communication error.The patient has not recharged the device in approximately 2.5 years.Consequently, the patient may undergo surgical intervention at a later date to address the issue.
 
Event Description
Follow-up information revealed the patient underwent surgical intervention on (b)(6) 2017 at which time the ipg was explanted and replaced.Reportedly, effective therapy was regained following the procedure and the issue is resolved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI IPG, 16-CHANNEL RECHARGEABLE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key6390043
MDR Text Key69439313
Report Number1627487-2017-01172
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/28/2015
Device Model Number3788
Device Lot Number4332665
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received04/13/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/07/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL 1194 (2), SCS ANCHOR; MODEL 3228, SCS ANCHOR
Patient Outcome(s) Other;
-
-