Customers reported to biomérieux that the stage ground strip (reference (b)(4)) was broke/missing on the vitek® ms instrument.An investigation was performed.Vitek® ms instrument configuration (.Cfg) files were returned to and reviewed by kratos (vitek® ms instrument manufacturer).There were no irregularities.The files are identical to the factory.Cfg file, for the five (5) files that were received and reviewed.A review of photos of the broke/missing stage ground strip and door, shows that the ground strip appears to be broken by binding and not cyclic fatigue, due to the breakage and abrupt bend in the stage ground strip.It appears it was pinched and this caused the breakage.Kratos conclusion is most likely a handling issue with the stage, either physically or via software stage move commands sent via putty (instrument serial communication application).The stage can also behave strangely and drive to the incorrect place, but this is usually accompanied by the configuration and stage location corrupting which would have alerted us to this cause.Kratos completed testing to confirm that other safety mitigations are correctly implemented on vitek® ms.Kratos confirmed there was no manufacturing issue or design change on the stage ground strip since 2011.There was only a re-draw of the same ground strip with a new design software in 2013.There was no cyclic metal fatigue as confirmed through kratos durability testing for the new door design.There was no loss or corrupt configuration files when reviewed by kratos.The dsp control board was replaced but it was not returned to check if it was defective.The above probable root causes could not be confirmed from the information and data received.The stage ground strip is a safety device located on the vitek ms® to prevent the user from electrical shock.When the instrument door is opened, the stage ground strip ensures contact between the instrument stage, the supporting the samples slides holder and the instrument ground.The risk of electrical shock is mentioned in the vitek® ms risk analysis, stated the severity is critical (per hazard definition), and the initial probability is occasional (possible for a relative few occurrences over the life of the product.) the final probability after mitigation by design is improbable (assumed not to occur or cannot be distinguished from zero).Kratos assessed the occurrence of the risk in case the stage ground strip as improbable (probability unchanged due to the multiple other failures which would be required to occur at exactly the same time).Kratos provided more details regarding the other mitigations in place to prevent the user or the field service engineer (fse) from electrical shock.Dc power supply is turned off: the software switches the instrument to standby.The sample plate is positioned behind the door opening and the sac is vented over a set period.Vacuum interlock: once the vacuum is close to atmospheric pressure then the hv interlock operates to shut off the hv voltages.Also when the vacuum is at atmospheric pressure then this would in effect provide a short circuit to the hv power supply and the power supply's in built protection circuits would then switch off the supply to protect the power supply (in practice this would happen well before atmospheric pressure is reached).So this a backup security for the vacuum interlock on the hv supply being off through vacuum pressure change.Grounding strip: as the plate carrier drives through the door then the contact spring makes a final physical short circuit and again the power supply's built in protection circuits would then switch off the supply to protect the power supply.Also, when the sample plate is then driven though the door, pushing it open, this contact ensures that a physical short circuit is made for any voltage on the sample plate.Again the power supply's built in protection circuits would then switch off the supply to protect the power supply.This represents a backup security of grounding strip to ensure hv is off (at zero volts) through physical grounding.Kratos stated that, although the risk of high voltage (hv) electrical shock is improbable when ground strip is faulty (ref impact analysis), the broken ground strip may lead, due to broken metal sharp edges in the vacuum, to hv breakdown electrical arcs in the instrument.Electrical arcs may lead to spectra bad quality, which will lead to no identification result.Additional maintenance of the instrument would then be needed due to damage of some components with electrical arcs (burnt marks to be cleaned for example).This risk on instrument damage will be added in a next revision of the kratos instrument service manual as follow : "possibility of hv electrical shock, if the ground strip is faulty.The vitek® ms instrument must not be used if the ground strip is not functioning correctly." changed to "possibility of hv electrical shock and hv breakdown that can cause damage to the instrument, if the ground strip is faulty.The vitek® ms instrument must not be used if the ground strip is not functioning correctly." a change request has been created for this update at biomérieux as well as kratos.The fse installed new replacement part on the vitek® ms instrument during preventative maintenance (pm).Records confirmed that the instrument was maintained and pm was completed according to procedures.Biomérieux will continue to monitor this issue for significant trends for one year and collect any relative information for potential events.
|