NIHON KOHDEN CORPORATION CNS-9700A; CENTRAL MONITOR SYSTEM Back to Search Results
Model Number CNS-9700A
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/17/2017
Event Type  malfunction  
Manufacturer Narrative
The customer reports that the cns (central nurses station) froze for about 30 minutes.Upon re boot, the cns received a no input error and would not boot into windows.The cns was switched out with a known working cns.The device was returned for evaluation and repair.The unit was cleaned and evaluated.The reported problem of the cns freezing and having no video was duplicated.The hard drive was replaced to resolve the issue.Repaired device was returned back to the customer.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
 
Event Description
The customer reports that the cns (central nurses station) froze for about 30 minutes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CNS-9700A
Type of Device
CENTRAL MONITOR SYSTEM
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, japan 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn: shama mooman
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgmt
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
2687488
MDR Report Key6393925
MDR Text Key69564345
Report Number8030229-2017-00065
Device Sequence Number1
Product Code MHX
UDI-Device Identifier04931921000663
UDI-Public04931921000663
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K023475
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/09/2017,01/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/09/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCNS-9700A
Device Catalogue NumberMU-971RA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/09/2017
Distributor Facility Aware Date01/17/2017
Device Age106 MO
Event Location Hospital
Date Report to Manufacturer03/09/2017
Date Manufacturer Received03/09/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/18/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-