Model Number 393-090
Device Problems Incorrect Measurement (1383); High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/21/2015
Event Type  malfunction  
Manufacturer Narrative
The radiometer field service engineer visited the hospital and collected the data logs for further analysis.The root cause of this problem is currently being investigated.The results of the investigation will be included in a follow-up report.This is a resubmission of the initial mdr, in which mfr report # was incorrectly stated as cfn-based rather than fei-based (3002807-2015-000034 rather than 3002807968-2015-000034).No fields, in addition to mfr report #, have been changed since the original submission.
 
Event Description
On (b)(6) 2015 a blood sample from a neonate patient was measured on the abl90 flex analyzer with the following result for glucose: 08:12h, 47 mmol/l.A sample from the patient was measured on the accuchek glucose meter straight after the analysis on the abl90 flex analyzer.The result of this measurement was 4.7 mmol/l based on the two results and the condition of the patient, the customer reported the measurement from the abl90 flex as false high.
 
Manufacturer Narrative
Not fei based: the mdr report for this case is not fei based for the initial and follow up report no.1.The mdr report no.Was missing the digits 968 and stated to be 3002807-2015-00034 instead of 3002807968-2015-00034.Was left blank in follow up 1 ((b)(6) 2016).
 
Manufacturer Narrative
Baed on the investigation of the data logs it has unfortunately not been possible to confirm the root cause of this case.The analyzer software has been upgraded to version 3.1 mr7 and no further complaints have been received from this customer.It has not been possible to get further information in this case, why on (b)(6) 2016 the case was closed in the radiometer complaint handling system.This is a resubmission of the follow-up no 1 mdr, in which mfr report # was incorrectly stated as cfn-based rather than fei-based (3002807-2015-000034 rather than 3002807968-2015-000034).No fields, in addition to mfr report have been changed since the original submission.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ABL90 FLEX
Type of Device
ABL90 FLEX GLUCOSE PARAMETER
Manufacturer (Section D)
RADIOMETER MEDICAL APS
aakandevej 21
broenshoej, broenshoej DK-27 00
DA  DK-2700
Manufacturer (Section G)
RADIOMETER MEDICAL APS
aakandevej 21
broenshoej, broenshoej DK-27 00
DA   DK-2700
Manufacturer Contact
kristina toft
aakandevej 21
broenshoej, broenshoej DK-27-00
DA   DK-2700
538273349
MDR Report Key6394420
MDR Text Key71410627
Report Number3002807968-2015-00034
Device Sequence Number1
Product Code CGA
Combination Product (y/n)N
Reporter Country CodeNZ
PMA/PMN Number
K120197
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 03/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other Health Care Professional
Device Model Number393-090
Device Catalogue Number393-090
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/04/2015
Initial Date FDA Received03/10/2017
Supplement Dates Manufacturer ReceivedNot provided
Not provided
Supplement Dates FDA Received03/29/2017
03/10/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age9 DA
-
-