• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER MEDICAL-KALAMAZOO TREATMENT RECLINER; CHAIR, EXAMINATION AND TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER MEDICAL-KALAMAZOO TREATMENT RECLINER; CHAIR, EXAMINATION AND TREATMENT Back to Search Results
Catalog Number 3500000710
Device Problem Split (2537)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/13/2017
Event Type  malfunction  
Manufacturer Narrative
Repair parts were sent to the customer and they made the repairs themselves.
 
Event Description
It was reported that that the cross bar in the back of the chair was split causing the chair to not support patient weight.No patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TREATMENT RECLINER
Type of Device
CHAIR, EXAMINATION AND TREATMENT
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
chanda burghard
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key6397090
MDR Text Key69961224
Report Number0001831750-2017-00065
Device Sequence Number1
Product Code FRK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/10/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number3500000710
Was Device Available for Evaluation? Yes
Date Manufacturer Received02/13/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/22/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-