The hill-rom technician found the brake caster, brake/steer caster and detent assembly needed to be replaced.Per the hill-rom service manual the affinity bed requires an effective maintenance program.We recommend that you perform semi-annual preventative maintenance.Check the tires for cuts, wear, tread life, etc.Apply the brake and check to ensure that the bed will not move (when the brake is activated).If the bed moves, inspect it for wear and adjust if required.See ¿caster assembly¿ on page 4-99.Apply the steering pedal and check the steering to ensure proper locking action when activated.See ¿caster assembly¿ on page 4-99.A search of the hill-rom maintenance records did not show hill-rom performed any preventative maintenance on this bed.It is unknown if the facility performs preventative maintenance on their beds.The technician replaced the brake caster, brake/steer caster and detent assembly to resolve the issue.Based on this information, no further action is required.
|