• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS CL; MONOFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS CL; MONOFOCAL IOLS Back to Search Results
Model Number Z9002
Device Problems Bent (1059); Break (1069); Crack (1135)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/06/2017
Event Type  malfunction  
Manufacturer Narrative
The lens was inserted and removed.(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Event Description
It was reported that an intraocular lens (iol) cracked upon insertion into the eye of a female patient.It was also reported bent/broken haptic and lens scratched.The lens was removed and a new lens was implanted during the same surgical procedure.No incision enlargement was reported.Reportedly, the patient has recovered.No further information was provided.
 
Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 03/21/201.7 device returned to manufacturer? yes.Device evaluation: the intraocular lens (iol) was returned at the manufacturing site for evaluation.Visual inspection at 10x microscope magnification showed fibers/particles and residue of viscoelastic on the lens which indicated that the unit was handled and prepare for surgical used.A cut was observed on the lens.One of the haptic was observed slightly distorted.Some scratches were also observed on the lens surface.The customer's reported complaint was verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no additional investigation requests for this order number have been received.Labeling review: the directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.The dfu states to examine the lens thoroughly to ensure particles have not become attached to it, and examine the lens optical surfaces for other defects.As a result of the investigation there is no indication of a product quality deficiency and the reported issue was verified.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS CL
Type of Device
MONOFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
Manufacturer (Section G)
ABBOTT MEDICAL OPTICS INC.
road 402 north, km 4.2
anasco industrial park, pob 14
anasco PR 00610
Manufacturer Contact
pam mcclain
1700 east st. andrew place
santa ana, CA 92705
7142478243
MDR Report Key6405072
MDR Text Key69949335
Report Number2648035-2017-00487
Device Sequence Number1
Product Code HQL
UDI-Device Identifier05050474530737
UDI-Public(01)05050474530737(17)210510
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P880081
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/15/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date05/10/2021
Device Model NumberZ9002
Device Catalogue NumberZ900200220
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/21/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/22/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/10/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CARTRIDGE PSCST30, LOT UNKNOWN
-
-