• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF) Back to Search Results
Model Number 97714
Device Problems Improper or Incorrect Procedure or Method (2017); Low Battery (2584); Device Displays Incorrect Message (2591); Device Operates Differently Than Expected (2913)
Patient Problems Stroke/CVA (1770); Seizures (2063); No Known Impact Or Consequence To Patient (2692)
Event Date 03/10/2017
Event Type  Injury  
Event Description
A patient reported via manufacturer representative a confirmed overdischarge event.The representative performed a 30 minute physician mode recharge (pmr) and then charged normally until he was able to clear the power on reset (por).The patient had a stroke or seizure on friday (b)(6) and a requested mri was unable to be performed due to the overdischarged battery.The representative cleared the por with 8840, but then the ins depleted below the charge level.No further complications were anticipated.The indication for implant was non-malignant pain.
 
Manufacturer Narrative
The event is no longer reportable for serious injury.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received from the representative (rep) and the consumer regarding the patient.It was reported that the patient did not charge her battery because she was in the hospital for an extended period of time.The rep performed a physician mode recharge and cleared the power on reset.As a result, the patient was able to recharge and gained stimulation.The rep reported that the seizure was due to medication she was on and not device related.No further complications were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
lisa clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key6409310
MDR Text Key70089401
Report Number3004209178-2017-05714
Device Sequence Number1
Product Code GZB
UDI-Device Identifier00643169109483
UDI-Public00643169109483
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 10/17/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/16/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/28/2016
Device Model Number97714
Device Catalogue Number97714
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received03/16/2017
Date Device Manufactured11/04/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age46 YR
-
-