MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION CNS-9700A; CENTRAL MONITOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION CNS-9700A; CENTRAL MONITOR SYSTEM Back to Search Results
Model Number CNS-9700A
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/30/2017
Event Type  malfunction  
Manufacturer Narrative
The biomedical engineer reports that the cns (central nurses station) will occasionally freeze and has to power cycle the device in order for it to work.Customer was sent a new hard disk drive.Replacing the hard drive resolved the issue.The defective hard drive was received, however due to it being unrepairable, the device has been discarded.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.Device unrepairable and discarded.
 
Event Description
The biomedical engineer reports that the cns (central nurses station) will occasionally freeze and has to power cycle the device in order for it to work.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CNS-9700A
Type of Device
CENTRAL MONITOR SYSTEM
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, japan 161-8 560
JA  161-8560
Manufacturer (Section G)
NIHON KOHDEN TOMIOKA CORPORATION
1-1 tajino
attn: shama mooman
tomioka city, gunma 370-2 314
JA   370-2314
Manufacturer Contact
shama mooman
safety mgmt dept, quality mgmt
seibu bldg 2, 4th floor 1-11-2
kusunokidai tokorozawa, saitama 359-8-580
JA   359-8580
2687488
MDR Report Key6412122
MDR Text Key70178698
Report Number8030229-2017-00074
Device Sequence Number1
Product Code MHX
UDI-Device Identifier04931921000663
UDI-Public04931921000663
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K023475
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 03/16/2017,01/30/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/16/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCNS-9700A
Device Catalogue NumberMU-971RA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/10/2017
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA03/16/2017
Distributor Facility Aware Date01/30/2017
Device Age124 MO
Event Location Hospital
Date Report to Manufacturer03/16/2017
Date Manufacturer Received03/16/2017
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/11/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-