MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 30°, STERILE, SINGLE USE; HF-RESECTION ELECTRODES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 30°, STERILE, SINGLE USE; HF-RESECTION ELECTRODES Back to Search Results
Model Number WA22306D
Device Problems Break (1069); Detachment Of Device Component (1104); Device Or Device Fragments Location Unknown (2590)
Patient Problem Foreign Body In Patient (2687)
Event Date 03/10/2017
Event Type  malfunction  
Manufacturer Narrative
The suspect medical devices have not yet been returned to the manufacturer for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical devices are returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
 
Event Description
Olympus was informed that during a combined diagnostic hysteroscopy and transcervical resection of the endometrium in saline (tcris) procedure, the ceramic insulation at the distal end of the resection sheath a22041a broke off and fell inside the patient's uterus.Furthermore, the loop wire at the distal end of the hf resection electrode wa22306d broke off as well.While the ceramic insulation was retrieved from the patient by unknown approach, the loop wire remained inside the endometrium since it could not be located by ultrasound.No further information was provided and it is unknown if and how the intended procedure was completed.However, there was no adverse event or patient injury.This is report 1 out of 2.
 
Manufacturer Narrative
Device evaluation: the suspect medical device was not returned to the manufacturer for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, a material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the hf resection electrode without showing any abnormalities.Furthermore, based on the customer description, the breakage of the loop wire was most likely caused by mechanical overload by the application of excessive force.The case will be closed from olympus side with no further actions.However, the reported phenomenon will be recorded for trending and surveillance purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 30°, STERILE, SINGLE USE
Type of Device
HF-RESECTION ELECTRODES
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer (Section G)
OLYMPUS MEDICAL PRODUCTS CZECH, SPOL. S R.O.
telickova 457/29
prerov
Manufacturer Contact
daniel wladow
kuehnstrasse 61
hamburg 
4940669662
MDR Report Key6412524
MDR Text Key70432781
Report Number9610773-2017-00043
Device Sequence Number1
Product Code FAS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK120418
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/21/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/17/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2021
Device Model NumberWA22306D
Device Catalogue NumberWA22306D
Device Lot NumberP16Y0008
Other Device ID Number14042761051689
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/20/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
HF UNIT "ESG-400" ((B)(6)); TCRIS HYSTEROSCOPE
Patient Age74 YR
-
-