OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 30°, STERILE, SINGLE USE; HF-RESECTION ELECTRODES
|
Back to Search Results |
|
Model Number WA22306D |
Device Problems
Break (1069); Detachment Of Device Component (1104); Device Or Device Fragments Location Unknown (2590)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 03/10/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
The suspect medical devices have not yet been returned to the manufacturer for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, if the suspect medical devices are returned for evaluation/investigation or additional significant information becomes available, this report will be updated.
|
|
Event Description
|
Olympus was informed that during a combined diagnostic hysteroscopy and transcervical resection of the endometrium in saline (tcris) procedure, the ceramic insulation at the distal end of the resection sheath a22041a broke off and fell inside the patient's uterus.Furthermore, the loop wire at the distal end of the hf resection electrode wa22306d broke off as well.While the ceramic insulation was retrieved from the patient by unknown approach, the loop wire remained inside the endometrium since it could not be located by ultrasound.No further information was provided and it is unknown if and how the intended procedure was completed.However, there was no adverse event or patient injury.This is report 1 out of 2.
|
|
Manufacturer Narrative
|
Device evaluation: the suspect medical device was not returned to the manufacturer for evaluation/investigation.Therefore, the exact cause of the user's experience and the reported phenomenon could not be determined and is being judged as unknown.However, a material or quality problem can be excluded since a manufacturing and quality control review was performed for the affected lot number of the hf resection electrode without showing any abnormalities.Furthermore, based on the customer description, the breakage of the loop wire was most likely caused by mechanical overload by the application of excessive force.The case will be closed from olympus side with no further actions.However, the reported phenomenon will be recorded for trending and surveillance purposes.
|
|