The event occurred during an iort treatment on thursday (b)(6) 2017 with liac sn (b)(4), installed in (b)(6) medical center - (b)(6), pertains to an interrupted treatment.The prescribed dose was 23.3 gy @ r100.The delivered dose has been 6 gy @ r100.During a technical analysis on liac sn (b)(4), carried out on installation site by the engineers of sit-sordina iort technologies spa on (b)(6) 2017, it has been detected that some connections of the wiring installed in the operating rooms were loose.Therefore the liac sn (b)(4) worked correctly causing the interruption of the treatment.
|