Model Number CD3367-40QC
Device Problems No Device Output (1435); Over-Sensing (1438)
Patient Problem Heart Failure (2206)
Event Date 02/15/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported when the patient was hospitalized due to the exacerbation of heart failure, noise was observed on the atrial and ventricular channels.The device was also noted at a state of no output.It is suspected the cause of the no output was an anomaly in the circuit of the device.The device was explanted and replaced.No adverse consequences were reported during and after the procedure.
 
Manufacturer Narrative
No conclusion code available.The reported field event of noise was confirmed in the laboratory due to review of the segms.The device was tested on the bench and noted there was an unstable hybrid supply.The cause of the unstable hybrid supply was found to be a capacitor connection anomaly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUADRA ASSURA ICD
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6426412
MDR Text Key70624843
Report Number2938836-2017-19203
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/02/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/22/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2016
Device Model NumberCD3367-40QC
Device Lot Number4878458
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received02/15/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/12/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age76 YR
-
-