(b)(4).Device evaluated by manufacturer: the device was returned for analysis.The device has no visible defects.Upon using the keyence/microscope there is visible evidence of 5 out of 6 of the irrigation ports are blocked from foreign material.There is blood in the tip which could cause the holes block.The ablation testing could not be verified by using the maestro generator 4000 and metriq pump, and the device was not found within specifications.Electrical test could not be conducted because of high impedance occurring from a block occlusions in the irrigation ports at the distal tip.No opens or shorts were found.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|
Reportable based on device analysis completed on 02-mar-2017.It was reported that noise was noted over the distal electrodes.A 7.5 110 2.5 blazer¿ open-irrigated was selected for use.During the procedure, when the catheter was inside the patient and after pacing the high right atrium, a baseline wonder type noise was noted over the distal electrodes.The bipol 1-2 pins were changed in the junction box and a new cable was tried but the problem persisted.The catheter was then removed and replaced with another of the same device and then only a mild noise was noted.No patient complications were reported and the patient's status was stable.However, device analysis revealed that 5 out of 6 of the irrigation ports were blocked with dried blood.
|