Model Number 9-ASD-014
Device Problem Unintended Movement (3026)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/01/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
Immediately after a 14mm amplatzer septal occluder (aso) was released from the delivery cable, it embolized into the descending aorta.The aso was snared and removed.A 20mm aso was successfully implanted.
 
Manufacturer Narrative
The results of this investigation confirmed the amplatzer septal occluder met all dimensional specifications when analyzed at sjm.A review of the device history record confirmed the occluder met all visual, dimensional, and functional specifications at the time it was manufactured, prior to shipment.There was no evidence to suggest there was an intrinsic defect in the occluder, and the cause for the embolization remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
CARDIAC OCCLUSION DEVICE
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6427173
MDR Text Key70633308
Report Number2135147-2017-00045
Device Sequence Number1
Product Code MLV
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/31/2020
Device Model Number9-ASD-014
Device Catalogue Number9-ASD-014
Device Lot Number4964377
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/28/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/07/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/02/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-