• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARIBEX NOMAD PRO 2; EXTRAORAL SOURCE X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARIBEX NOMAD PRO 2; EXTRAORAL SOURCE X-RAY SYSTEM Back to Search Results
Model Number 0.850.0022
Device Problem Battery Problem (2885)
Patient Problem No Patient Involvement (2645)
Event Date 01/19/2017
Event Type  malfunction  
Manufacturer Narrative
Upon initial receipt of the event, it was determined that the event was not reportable based on the customer information provided.An engineering review on 3/15/2017 determined that this event is reportable.The unit was returned and an engineering evaluation has been completed.Upon visual inspection, it appears that a thermal event occurred.It is not possible to determine the exact sequence of events that led to the thermal event.However, there is evidence that a short occurred in the lower battery pack between the cells and pcb.There was a blackening on the top of the lower battery pack and its printed circuit board (pcb).Cells 2 and 3 appear to have been involved in the thermal event, likely by supplying energy into a fault in the board.Additionally there was a slight melting of the handset plastic enclosure.This concludes the investigation.A recall is ongoing.Reference z-2716/2717-2016.
 
Event Description
It was reported that the battery was not charging.There was no patient or user involvement reported and no injuries reported.The unit was not in use at the time of the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NOMAD PRO 2
Type of Device
EXTRAORAL SOURCE X-RAY SYSTEM
Manufacturer (Section D)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer (Section G)
ARIBEX
11727 fruehauf drive
charlotte NC 28273
Manufacturer Contact
erika martin
11727 fruehauf drive
charlotte, NC 28273
7045877241
MDR Report Key6432839
MDR Text Key70902983
Report Number1017522-2017-00025
Device Sequence Number1
Product Code EHD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K081664
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Dentist
Type of Report Initial
Report Date 01/19/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0.850.0022
Device Catalogue Number0.850.0022
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/07/2017
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/19/2017
Initial Date FDA Received03/24/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/29/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-2716/2717-2016
Patient Sequence Number1
-
-