• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. MCK TIBIAL BASEPLATE-LM/RL-SZ 5; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. MCK TIBIAL BASEPLATE-LM/RL-SZ 5; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM Back to Search Results
Catalog Number 180605
Device Problem Loss of Osseointegration (2408)
Patient Problems Discomfort (2330); Injury (2348); Inadequate Osseointegration (2646)
Event Date 09/19/2013
Event Type  Injury  
Manufacturer Narrative
An evaluation of the device cannot be performed as the device was retained by the hospital and was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Event Description
Tibial loosening with mechanical failure, left uka.Revision of tibial compartment.
 
Manufacturer Narrative
An event regarding loosening involving a mako baseplate was reported.The event was not confirmed.Method & results: -device evaluation and results: a visual, functional and dimensional inspection could not be performed as the device was not returned.-medical records received and evaluation: the provided medical information was submitted to a consulting clinician who indicated: "there are no x-rays available for review, no examination of the explanted components, no confirmation of a loose primary tibial component, and no follow-up status-post revision with a different uka femoral component subsequent to (b)(6) 2013.There is no indication that factors of implant component design, manufacturing or materials were responsible for this clinical situation." -device history review: device history review indicated the devices accepted into final stock from the reported lot were free from discrepancies -complaint history review: there has been no other event for the lot referenced.Conclusions: the event could not be confirmed nor the root cause determined because the devices were not returned for evaluation and insufficient information was provided.Provided medical records were submitted to a consulting clinician who indicated that no confirmation of a loose primary tibial component and there is no indication that factors of implant component design, manufacturing or materials were responsible for this clinical situation.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
Tibial loosening with mechanical failure, left uka.Revision of tibial compartment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCK TIBIAL BASEPLATE-LM/RL-SZ 5
Type of Device
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
keyla colon
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key6439379
MDR Text Key71033822
Report Number3005985723-2017-00152
Device Sequence Number1
Product Code NPJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090763
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type study
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/31/2015
Device Catalogue Number180605
Device Lot Number26020910
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/27/2017
Initial Date FDA Received03/28/2017
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/25/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/04/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age71 YR
Patient Weight87
-
-